Pharma Chambers
Pharma Climatic Test Chambers, Series CP
For long-term stability testing with highest testing precision.
Standard version of the product line “CP”:
- Capacitive humidity measuring system
- Casing in stainless-steel
- Door lockable
- Entry port ø 50 mm
- 6 shelves
- CTS standard -terminal
- Low noise refrigerating unit
- Potential-free contact for malfunction signal
- Interface RS 232
CTS user benefits:
- Use of environmentally friendly materials and refrigerants
- Own CTS-controller optimised to CTS-systems
- Simple user-friendly operation
- Internal and external material in stainless-steel for easy cleaning
Options:
- Door with window
- Validation of the CTS unit (IQ, OQ)
- Software validation
- Test space capacity in liters:
- 500 / 1100
Temperature range:
+10 °C / +50 °C
Humidity range:
10 – 90 % RH
Dew Point range I:
+7 °C / +45 °C
Dew Point range II:
-5 °C / +7 °C
Design:
Control: microprocessor control and monitoring system
Test space: stainless steel grade 1.4301
Casing: stainless steel grade 1.4301
Refrigerants: chloride free, hermetically sealed
Humidity system: capacitive measuring
Options:
- add. entry ports ø 50 mm
- add. shelves
- add. Pt 100 for temperature measuring on test specimen alternative reversible to controller function
- door with full-size window (reduced dew-point-range < +30 °C)
- extended temperature range down to ± 0 °C
- reservoir for humidification water
- temperature protection for test-specimen
- Touch Panel, colour display LCD
- CID software for programming and documentation
- RS 232 interface changeable to USB
| Type |
CP 10/500 |
CP 10/1100 |
||
| Test Space |
liters |
500 |
1100 |
|
| Test Space Dimensions | height |
mm, appr. |
1225 |
1225 |
| width |
mm, appr. |
595 |
2 x 595 = 1190 |
|
| depth |
mm, appr. |
620 |
620 |
|
| Overall Dimensions | height |
mm, appr. |
2080 |
2080 |
| width |
mm, appr. |
720 |
2 x 720 = 1440 |
|
| depth |
mm, appr. |
850 |
850 |
|
| Temperature Tests | ||||
| Temperature Range |
°C |
+10 / +50 |
||
| Temperature Fluctuation |
K |
± 0,5 temporally |
||
| Climatic Tests | ||||
| Temperature Range |
°C |
+10 / +50 |
||
| Humidity Range |
% RH |
10 / 90 |
||
| Dew Point Range I |
°C |
+7 / +45 |
||
| Dew Point Range II |
°C |
-5 / +7 |
||
| Humidity Fluctuation |
% RH |
± 1 / 3 |
||
| Temperature Fluctuation |
K |
± 0,5 temporally |
||
| Nominal Voltage |
230 V +6% -10%, 1/N, 50 Hz |
|||
| Nominal Power |
kW, appr. |
1,6 |
2,0 |
|
| Condenser |
air-cooled |
|||
All figures are average values which have been obtained at a temperature of +25° C, without test specimens, without thermal load and without options.
We reserve the right to make alterations due to technical development.